Site icon Rudolph Research Analytical

Especificaciones del sacarímetro

Automatic Saccharimeter Model Comparisons

Rudolph makes it easy to match an Automatic Digital Saccharimeter to your budget and industry application. The Autopol IZ and IIZ were designed for limited budget applications where 0.02°Z (ISS) accuracy is sufficient. For the more demanding accuracy applications, Rudolph offers the Autopol 589, 880, 880 Plus and 880T models that have ±0.01°Z (ISS) accuracy, OD 4.0 dark sample penetration power, and feature-rich software.

Especificaciones

Autopol® IZ Saccharimeter Autopol® IIZ Saccharimeter Autopol® 589 Saccharimeter
 Optical Wavelengths  589nm  589nm, 880nm 589nm
 Wavelength Selection  Fijo  Selectable, 2 only  Fijo
 Measuring Mode  Optical Rotation, Specific Rotation, Concentration, Sugar Degrees, °Z (ISS), User Defined Scale Optical Rotation, Specific Rotation, Concentration, Sugar Degrees, °Z (ISS), User Defined Scale °Z (ISS), Optical Rotation or User Defined Scale
 Measuring Range  ±89.9° Arc Optical Rotation, ±259 °Z (ISS)  ±89.9° Arc Optical Rotation, ±259 °Z (ISS)  ±89.9° Arc Optical Rotation, ±259 °Z (ISS)
 Resolution  0.01° Arc Optical Rotation, 0.01 °Z (ISS) 0.01° Arc Optical Rotation, 0.01 °Z (ISS) 0.001° Arc Optical Rotation, 0.01 °Z (ISS)
 Reproducibility  0.01° Arc Optical Rotation, 0.01 °Z (ISS)  0.01° Arc Optical Rotation, 0.01 °Z (ISS)  0.001° Arc Optical Rotation, 0.01 °Z (ISS)
 Accuracy  0.02° Z  0.02°Z (ISS) Sugar Degrees 0.01 °Z (ISS)
 Measuring Scale  Degrees Arc, Optical Rotation, °Z (ISS)  Degrees Arc, Optical Rotation, °Z (ISS)  Degrees Arc, Optical Rotation, °Z (ISS)
 Prism  Calcita Glan Thompson  Calcita Glan Thompson Calcita Glan Thompson
 Temperature Control  By external Water Bath  By external Water Bath  By external Water Bath
 Temperature Correction  18°C – 40°C  18°C – 40°C  18°C – 40°C
 TempTrol™ Accuracy  Temperature accuracy determined by Water Bath  Temperature accuracy determined by Water Bath Temperature accuracy determined by Water Bath
 Temp. Probe Range  10° – 40°C  10° – 40°C 10° – 40°C
 Temp. Probe Accuracy ±0.1°C ±0.1°C ±0.1°C
 Measurement Time  5 mediciones en menos de 25 segundos (media) 5 mediciones en menos de 25 segundos (media) 12° per second
 Light Source Standard  Tungsten-Halogen 6V, 20W, avg. 2,000 hour life  Tungsten-Halogen 6V, 20W, avg. 2,000 hour life Tungsten-Halogen 6V, 20W, avg. 2,000 hour life
 Light Source Optional  LED 20-30 mA 100.000 hr. vida media  NA LED 20-30 mA 100.000 hr. vida media
 Sample Chamber  Accepts sample tubes up to 200 mm  Accepts sample tubes up to 200 mm  Accepts sample tubes up to 200 mm
 Data Storage  32GB SSD  32GB SSD  32GB SSD
 Communication Interface

1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports

1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports 1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports
 Calibration  Calibrado automático mediante pantalla táctil  Calibrado automático mediante pantalla táctil  Calibrado automático mediante pantalla táctil
 Purity Software  Optional with PC  Optional with PC  Estándar
 Connection to Refractometer  Estándar  Estándar  Estándar
 Display  7.5 cm x 10 cm Graphics LCD, 320 x 240 dots cold fluorescent back lit  7.5 cm x 10 cm Graphics LCD, 320 x 240 dots cold fluorescent back lit  12.0 cm x 16.0 cm Color TFT Graphics LCD with LED backlight providing 800 x 600 resolution with 600 nits of brightness
 User Interface  Pantalla táctil  Pantalla táctil  Pantalla táctil
 Automatic Sensitivity Control Measures samples with transmittance as low as 0.1% (up to OD 3.0) Measures samples with transmittance as low as 0.1% (up to OD 3.0) Mide muestras con una transmitancia tan baja como 0,01% (hasta OD 4,0)
 Input Power  100 – 240V, 50/60 Hz  100 – 240V, 50/60 Hz  100 – 240V, 50/60 Hz
 Operating Dimensions  24.3” W x 12.7” H x 17.5” D (617mm W x 323mm H x 445mm D) 24.3” W x 12.7” H x 17.5” D (617mm W x 323mm H x 445mm D)  35” W x 10.5” H x 17” D (890mm W x 267mm H x 432mm D)
 Shipping Dimensions  31” W x 28” H x 26” D (787.4mm W x 711.2mm H x 660.4mm D) 31” W x 28” H x 26” D (787.4mm W x 711.2mm H x 660.4mm D) 42” W x 24” H x 21” D (1066.8mm W x 609.6mm H x 533.4mm D)
 Operating Weight  42 libras (19,05 kg)  42 libras (19,05 kg)  85 lbs. (39 kg)
 Shipping Weight  32,6 kg (72 libras) 32,6 kg (72 libras)  120 lbs. (54.5 kg)

Especificaciones

Autopol® 880/880T Autopol® 880T/880 Plus Saccharimeter
 Optical Wavelengths Standard: 589nm, 880nm
Optional: 587nm, 882nm
Standard: 589nm, 880nm
Optional: 587nm, 882nm
 Wavelength Selection  Selectable: 2 standard, 4 optional Selectable: 2 standard, 4 optional
 Measuring Mode °Z (ISS), Optical Rotation or User Defined Scale °Z (ISS), Optical Rotation or User Defined Scale
 Measuring Range ±89.9° Arc Optical Rotation, ±259 °Z (ISS)  ±89.9° Arc Optical Rotation, ±259 °Z (ISS)
 Resolution 0.001° Arc Optical Rotation, 0.01 °Z (ISS)  0.001° Arc Optical Rotation, 0.01 °Z (ISS)
 Reproducibility 0.001° Arc Optical Rotation, 0.01 °Z (ISS)  0.001° Arc Optical Rotation, 0.01 °Z (ISS)
 Accuracy 0.01 °Z (ISS)  0.01 °Z (ISS)
 Measuring Scale  Degrees Arc, Optical Rotation, °Z (ISS)  Degrees Arc, Optical Rotation, °Z (ISS)
 Prism  Calcita Glan Thompson  Calcita Glan Thompson
 Temperature Control  Standard: By external Water Bath
Optional: TempTrol™ electronic heating & cooling 18°C – 35°C
 TempTrol™ Automatic Electronic Heating & Cooling 18°-35°C (standard) Autopol 880T Only
 Temperature Correction  18°C – 40°C  18°C – 40°C
 TempTrol™ Accuracy Temperature accuracy determined by external Water Bath  ±0.2°C TempTrol accuracy Autopol 880T Only
 Temp. Probe Range  10° – 40°C  10° – 40°C
 Temp. Probe Accuracy  ±0.1°C  ±0.1°C
 Measurement Time 12° per second  12° per second
 Light Source Standard Tungsten-Halogen 6V, 20W, avg. 2,000 hour life  Tungsten-Halogen 6V, 20W, avg. 2,000 hour life
 Light Source Optional LED 20-30 mA 100.000 hr. vida media LED 20-30 mA 100.000 hr. vida media
 Sample Chamber  Accepts sample tubes up to 200 mm  Accepts sample tubes up to 200 mm
 Data Storage  32GB SSD  32GB SSD
 Communication Interface 1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports 1 x Serial, 1 x VGA, 1 x HDMI, 2 x Cat 5, 4 x USB ports
 Calibration  Calibrado automático mediante pantalla táctil  Calibrado automático mediante pantalla táctil
 Purity Software  Estándar  Estándar
 Connection to Refractometer  Estándar  Estándar
 Display  12.0 cm x 16.0 cm Color TFT Graphics LCD with LED backlight providing 800 x 600 resolution with 600 nits of brightness  10.4” diagonal, 800- 600 pixels, color, Flat Panel Monitor with resistant Touch Screen Interface, 600 nits brightness, gasketted for spill protection
 User Interface  Pantalla táctil  Pantalla táctil
 Automatic Sensitivity Control  Mide muestras con una transmitancia tan baja como 0,01% (hasta OD 4,0)  Mide muestras con una transmitancia tan baja como 0,01% (hasta OD 4,0)
 Input Power  100 – 240V, 50/60 Hz  100 – 240V, 50/60 Hz
 Operating Dimensions  35” W x 10.5” H x 17” D (890mm W x 267mm H x 432mm D)  32” W x 11.5” H x 18” D (813mm W x 292mm H x 457mm D)
 Shipping Dimensions  42” W x 24” H x 21” D (1066.8mm W x 609.6mm H x 533.4mm D)  43” W x 28” H x 23” D (1411mm W x 919mm H x 722mm D)
 Operating Weight 85 lbs. (39 kg)  95 lbs. (43 kg)
 Shipping Weight 120 lbs. (54.5 kg)  130 lbs. (59 kg)
Exit mobile version