Polarimeter – Quartz Plate Calibration Standards
Rudolph Research endeavors to adhere to the highest know calibration standards in the laboratory instrument industry. To view our full NVLAP (ISO/IEC 17025:2017) certificate and Scope of Accreditation, please visit our Quality page.
Manufactured to NIST and ICUMSA Standards
Rudolph Research Analytical Quartz Control Plates provide a permanent calibration standard for polarimeters and saccharimeters. Every Rudolph calibration standard is manufactured using extremely stable crystalline quartz which is manufactured and polished to specific planeness, parallelism, purity, and axis standards. These standards meet or exceed the specifications outlined on Page 61 of the National Institute for Standards and Technology (NIST) Circular C440 or as laid down by the International Commission for Uniform Methods of Sugar Analysis (ICUMSA). Each Quartz Plate comes complete with a label showing calibration at 11 different wavelengths (Picture G), a protective box and a NIST Traceable Certificate.
Selecting a Quartz Calibration Standard
Selecting a Quartz Plate
Any rotation is available but generally available stock rotations are listed below. Quartz Plates come in single and dual versions. Single Quartz Plates have a single rotation value at each wavelength and are calibrated in °Arc (Optical Rotation) and °ISS/°Z (Sugar Degrees). Dual Plates are also calibrated in °Arc and °ISS, but have three rotation values at each wavelength. This is possible because each plate is made up of one left or levo turning Quartz Plate (i.e. -10° Arc) and one right or dextro turning plate (i.e. +12° Arc) which may be used separately or together for a combined rotation of 2° Arc (+12°-10°). Every Rudolph Research Analytical Quartz Plate is manufactured to a reported uncertainty of ±0.002° Arc at 589. Please contact the factory for the Quartz Accessory Plate that is best suited for your needs.
Call Us To Select A Calibration Standard That Meets Your Application
TempTrol™ Quartz Plates indicated by the letter “T” have the facility for automatic temperature control when purchased with an Autopol® IV or V with TempTrol.
TempTrol Quartz Control Plate A700T-1
Standard Quartz Control Plate A700-1
Has a rotation between +10 to +35 angular degrees (°Arc) or +28 to +102 International Sugar Scale degrees (°ISS)/(°Z). The Quartz Control Plate consists of one right turning quartz plate.
Standard Rotations at 589 nm: 10.5°, 11°, 11.5°, 12°, 12.5°, 13°, 15°, 16°, 20°, 21.5°, 23°, 34°, 34.5°
Accuracy at 589 nm: ±0.002° Arc
TempTrol Quartz Control Plate 300T-S
Standard Quartz Control Plate 300-S
Has a rotation between +10 to +35 angular degrees (°Arc) or +28 to +102 International Sugar Scale degrees (°ISS)/(°Z). The Quartz control plate consists of one right turning quartz plate which is to be within ±0.5° Arc or ±1.5° ISS of the customer’s specified rotation.
Rotation: Customer Selected
Accuracy at 589 nm: ±0.002° Arc
TempTrol Quartz Control Plate A700T-2
Standard Quartz control plate A700-2
Has a combined rotation between +1 to +10 angular degrees (°Arc) or +2 to +30 International Sugar Scale degrees (°ISS)/(°Z). The Quartz Control Plate consists of one left and one right turning quartz plate in separate cells, each of which can be removed from the cell holder for using either plate separately or together for a combined rotation.
Standard Rotations at 589 nm: approx. 1°, 2°, 6° (See Picture H)
Accuracy at 589 nm: ±0.002° Arc
TempTrol Quartz Control Plate 301T-S
Standard Quartz Control Plate 301-S
Has a combined rotation between +1 to +10 angular degrees (°Arc) or +2 to +30 International Sugar Scale degreez (°ISS)/(°Z). The Quartz Control Plate consists of one left and one right turning quartz plate in separate cells, each of which can be removed from the cell holder for using either plate separately or together for a combined rotation. Quartz Control Plates are to be within ±0.5° Arc or ±1.5°ISS of the customer’s specified rotation.
Rotation: Customer Selected
Accuracy at 589 nm: ±0.002° Arc
TempTrol Quartz Control Plate 302T
Standard Quartz Control Plate 302
Has a rotation between -10 to -35 angular degrees (°Arc) or -28 to -102 International Sugar Scale degrees (°ISS)/(°Z). The Quartz Control Plate consists of one left turning quartz plate.
Standard Rotations at 589 nm: -10°, -10.5°, -10.75°, -11°, -11.25°, -12°, -15°, -16°, -22°
Accuracy at 589 nm: ±0.002° Arc
TempTrol Quartz Control Plate 302T-S
Standard Quartz Control Plate 302-S
Has a combined rotation between -1 to -10 angular degrees (°Arc) or -2 to -30 International Sugar Scale degrees (°ISS)/(°Z). The Quartz Control Plate consists of one left and one right turning quartz plate in separate cells, each of which can be removed from the cell holder for using either plate separately or together for a combined rotation. Quartz Control Plates are to be within ±0.5° Arc or ±1.5°ISS of the customer’s specified rotation.
Rotation: Customer Selected
Accuracy at 589 nm: ±0.002° Arc
Recalibration of Quartz Control Plates
Neither NIST nor ICUMSA provide guidelines for the frequency of Polarimeter Quartz Control Plate Recalibration. Since quartz is optically stable over time, changes in its optical rotation are the result of optical stress or surface scratches caused by physical handling. Therefore, it is the opinion of Rudolph Research Analytical that recalibration of a Quartz Calibration Standard every two to three years is satisfactory. For further assistance with quartz accessory plate recalibration, send an email to info@rudolphresearch.com, or call our factory at +1-973-584-1558.
Every Quartz Control Plate is calibrated by Rudolph Research Analytical and is traceable to the National Institute of Standards and Technology (NIST).