Laboratory Accuracy Everywhere You Need to Measure
Unmatched Capabilities in a Fully Portable Density Meter
Temperature Controlled Measurements at 20° and 25°C
The problem with most handheld Density Meters is they drift because they do not control temperature. Density Meter measurements are highly sensitive to temperature and mathematically correcting for temperature may not be good enough for some applications. Only temperature control will result in the accuracy many users require.
The Rudolph DDM 29 utilizes powerful and compact Peltier technology. Control of the sample temperature for precise heating and cooling to ensure stable and accurate measurements at either 20 or 25°C.
Rugged Stainless Steel Double U-Tube for unmatched accuracy, durability, and weight reduction
The Rudolph DDM29 uses a 316 Stainless Steel, Double U-Tube instead of a simple glass U-Tube. This new design is durable and corrosion resistant. Stainless Steel is preferable for customers who are tired of replacing glass U-Tubes in other handheld brands.
The Double U-Tube design allows for handheld accuracy and rapid measurement speed. Demonstrating 3 decimal place measurement readings in 30 seconds, and 4 decimal measurements in 60 seconds.
User friendly GUI and Bright, Back Lit,Touch Screen Display
Extremely easy to use Graphic User Interface (GUI) and the availability of 10 different Languages on a brightly lit 7 inch(128mm) touch screen makes this the most User Friendly Density Meter in the market today. The interface is a bright LED touchscreen, with a large easy to navigate GUI. No more difficulties reading measurements in bright or dark environments.
Portability and Flexibility
TheDDM29 is powered by a Lithium-Ion battery that is easily recharged, your DDM29 is ready to go everywhere you need to make measurements. This might be at the loading dock when receiving incoming raw goods or out on a distillery floor. The Lithium Ion battery can be charged in approximately 2 hours providing up to 4 hours of constant measurements under heavy usage or 8 hours of normal usage. Included is a Charging Station for convenient stationary charging.
Laboratory Accuracy – Everywhere You Need to Measure
Most Handheld Density Meters use temperature correction and publish 2 or 3 digit accuracy which is very prone to drift. The DDM29 uses Peltier Temperature Control, which eliminates drift and provides accuracy to the 4th decimal place. Laboratory accuracy in the DDM29 reduces testing of materials in the main Lab. Testing can now be done accurately in the receiving dock, formulations area, or where ever measurements are needed.
Flexible Method Management
Factory installed measurement methods cover a wide group of industries and applications. The DDM29 is factory configured with over 50 density and concentration tables, such as Brix, %Alcohol, API, and most common chemicals. Additional tables, polynomials, and formulas may be added.
- Essential Oils
- Chemicals
- Beer, wort, and fermentation monitoring
- Raw material soft drink monitoring
- Monomers, Polymers and Elastometers
- Pharmaceutical raw materials and finished products
- Wort, juice, fermentation, spirits safe monitoring
- Blending, bottling, spirits and liquors packaging
- Sulfuric Acid in lead-acid batteries
- Adhesives, glues
- Agriculture, chemicals and fertilizers
- Colloids, nanotechnology
- Surfactants, detergents
- Emulsions
- Paints, Inks, Toners
- Organic, inorganic chemicals
- Fuel cells, power generation, sustainable fuels
- Soft drinks, juices, tea, coffee
- Petroleum samples according to ASTM methods
A measurement method is created exactly for unique measurement applications. You may use Concentration Tables, Formulas
and Polynomials to match the measurement methods used in your laboratory or field application.
Industry Applications For Handheld Density Meters
Technical Specifications
DDM29 Handheld Density Meter
Accuracy | Density: 0.0005 g/cm3* | Temperature: ±0.1°C |
Repeatability (Standard Deviation) | Density: 0.0001 g/cm3 Temperature: 0.02 °C |
Resolution | Density: 0.0001 g/cm3 Temperature: 0.01°C |
Temperature Selectable | 20 or 25 °C ±0.01°C |
Temperature Control (Controlled via Peltier) | 20 or 25°C. Selectable up to 0.001°C |
Ambient Temperature | -10 to + 55°C |
% Concentration Scales | Unlimited |
Ambient Temperature Limit | 0 to 35°C |
Pressure Range | 0 to 10 bar (145 psi ) |
Measurement Modes | Continuous, Single, Multiple |
Measurement Technique | Mechanical Oscillating Double U-Tube Method |
Measurement Range | Density: 0-3g / cm3 |
Minimum Sample Volume | Approx 1mL |
Wetted Materials | 316 Stainless steel, Teflon PTFE, ECTFE, PVDF |
Supported scales | Density, Specific Gravity, Alcohol Tables, Sugar/Extract, API functions, H2SO4 tables, acids, bases, organic and inorganic chemicals |
Operating System | Android OS |
Measurement Time | Typically 30 – 40 seconds after thermal equilibration |
Display | Bright LED – 7” touch screen. |
Communication Interfaces | USB-C, WiFi, Bluetooth® –Manual Entry, RFID, and Barcode for entering sample IDs are all Standard |
Printing | Wireless Printing Via Airprint |
Remote Support | Troubleshooting, Diagnostics Tools, Software Updates available via the Internet |
Internal Memory | Up to 3,000 Measurements |
Operating Dimensions | 7.28” (L) x 4.83” (W) x 13.59” (H) 18.49 cm (L) x 12.27 cm (W) x 34.52 cm (H) |
Shipping Dimensions | 18” (L) x 18” (W) x 12” (H) |
Operating Weight | 3lbs 6oz |
Charging /Power Consumption | 2 hours charging provides up to 4 hours of measurements in heavy usage or 8 hours under normal usage. |
Included with Instrument | Charging station, syringes, instructions |
Protection Class | IP54 |
Country of Manufacture and Design / Warranty | 1 Year Domestic and international warranty |